CONTACT US
Sales@hplc.com
Ph: 800-842-4752
Fx: 540-667-6878


Calibration Curve for Saccharides and Organic Acids (SH1011)

Quantitaive analysis of saccharides and organic acids was performed using SUGAR SH1011. The linearity of the calibration curves are fairly good.


Sample : 10micro-L
Correlation Coefficient
(0.2 to 250micro-g)
Glucose 0.9999
Succinic acid 1.0000
Isomaltotriose 0.9999
Xylitol 0.9999
Maltitol 1.0000
Acetic acid 1.0000
Propionic acid 1.0000
Adipic acid 0.9997


Column       : Shodex SUGAR SH1011 (8.0mmID*300mm)
Eluent       : 5mM H2SO4 aq.
Flow rate    : 0.6mL/min
Detector     : Shodex RI        for saccharides
               UV(210nm) for organic acids
Column temp. : 60deg-C