Margarine
Vitamin A palmitate was analyzed using GPC 
KF-801. Wave 
length of 325nm at which it has maximum UV absorption was used for the detection. 
Usually, saponification is necessary to analyze vitamin A as the sample pretreatment. 
However, in this analysis, no complicated sample pretreatment procedure was necessary. 
What is necessary is just to dissolve the sample in the eluent(THF) and filtrate 
it using a 0.45 micron disposable filter. This method using KF-801 is a very simple 
way to analyze vitamin A. 

Sample : 
Margarine 0.2%, 100micro-L
1. 
Vitamin A palmitate
Columns      : Shodex GPC KF-801 (8.0mmID*300mm) x 2
Eluent       : THF
Flow rate    : 1.0mL/min
Detector     : UV(325nm)
Column temp. : 40deg-C