Poly(-N,N-Dimethyl Aminomethyl Styrene) (1) (DMF Eluent)
Polymers, whose polarity is lower than that of DMF are adsorbed by the packing 
materials of the GPC 
KD-800 columns. For such polymers, 
the use of the OHpak 
SB-800 HQ columns is recommended because 
their packing material has a weaker hydrophobic interaction than that of the KD-800 
columns. Generally, in DMF eluent some salts like LiBr should be added for the 
SEC analysis of polymers having ionic groups. Salt is added to prevent ionic interaction 
between sample and packing material. 
 
  
  Sample : Poly(-N,N-dimethyl aminomethyl styrene)
Columns      : Shodex GPC KD-804 +KD-802.5, OHpak SB-805 HQ + SB-802.5 HQ (8.0mmID*300mm each)
Eluent       : 5mM LiBr in DMF
Flow rate    : 1.0mL/min
Detector     : Shodex RI
Column temp. : 40deg-C