Rubber (Toluene Eluent)
      
      Natural rubber was analyzed using GPC 
KF-802 and
      
KF-806M (columns for organic SEC (GPC)). Since
      natural rubber is easily dissolved in toluene, toluene eluent is used for
      GPC analysis of this sample.
      
      

      Sample : 
Rubber
      
      Columns      : Shodex GPC KF-806M x 2 + KF-802 (8.0mmID*300mm each)
Eluent       : Toluene
Flow rate    : 1.0mL/min
Detector     : Shodex RI
Column temp. : Room temp.