Analysis of Antioxidant in Polystyrene (GPC/MS)
Antioxidant (Irganox1010) in Polystyrene was analyzed with GPC/MS using GPC 
KF-402HQ (a semi-micro organic SEC (GPC) column). 
Only a low molecule weight domain can be poured into a mass spectrometer (MS) by using GPC mode. Therefore, it is not necessary to pretreat.
Please refer to 
Simultaneous Analysis of Various Additives of 
Polymer in Polystyrene (GPC/MS)

Sample : 1ppm Irganox 1010 in 1000ppm Polystyrene, 5micro-L
1. Polystyren
2. 
Irganox 1010
(HPLC)
Instrument   : Agilent 1100
Column       : Shodex GPC KF-402HQ (4.6mmID*250mm) x 2
Eluent       : THF
Flow rate    : 0.3mL/min
Detector     : UV(210nm)
Column temp. : 40deg-C
(MS)
Instrument      : Thermo Quest LCQ Advantage
Ionization      : APCI
Capillary temp. : 180deg-C
Vaporizer temp. : 450deg-C
Source voltage  : 6000V
Sheath gas flow : 80L/min