Comparison of KF-602 and KF-802
Downsized organic SEC (GPC) Column, GPC 
KF-602 (6.0mmID*150mm) and standard 
type column
 KF-802 (8.0mmID*300mm) are comapred using styrene 
oligomer. The required analysis time is 10 min. for KF-602 and 20 min. for KF-802. 
This means the analysis time for KF-602 is about half of that for KF-802. The 
solvent consumtion is 6mL (0.6mL/min x 10min.) for KF-602 and 20mL (1.0mL/min 
x 20min.) for KF-802. This means the solvent consumption for KF-602 is less than 
one third of that for KF-802. 

Sample: 
Styrene oligomer
              (Above)                                  (Below)
Columns      : Shodex GPC KF-602 (6.0mmID*150mm) x 2  : Shodex GPC KF-802 (8.0mmID*300mm) x 2
Eluent       : THF                                    : THF
Flow rate    : 0.6mL/min                              : 1.0mL/min
Detector     : Shodex RI(small cell volume)           : Shodex RI
Column temp. : 40deg-C                                : 40deg-C