Effect of Injection Volume
      
      For GPC 
KF-606M, the effect of injection volume
      on elution volume and peak-width is shown here. When the injection volume
      is changed, the elution volume changes very little, however, the
      peak-width inceases markedly when the injection volume exceeds 20 micro-L.
      Therefore, the use of sample concentration lower than 0.05% and the
      injection volume smaller than 20 micro-L is recommended. 
      
      

      
      Sample : 
PS, Polystyrene
      (Elution Volume)
      1.7K(0.05%), 4.8K(0.05%), 28.5K(0.05%), 66K(0.05%), 156K(0.03%),
      514K(0.03%), 2950K(0.01%), 6750K(0.01%)
      (Peak-width)
      1.7K(0.05%), 4.8K(0.05%), 28.5K(0.05%), 66K(0.05%), 156K(0.01%),
      514K(0.01%), 2950K(0.01%)
      
Column       : Shodex GPC KF-606M (6.0mmID*150mm)
Eluent       : THF 
Column temp. : 40deg-C