MW Distribution Analysis (KF-406L HQ)
MW distribution of SRM 706 was measured using organic solvent SEC (GPC) semi-micro column, GPC 
KF-406LHQ and conventional organic solvent SEC (GPC) column, GPC 
KF-806M. Both columns provided good agreements even using different detector systems (Light scattering and sedimentation equilibrium) that provide absolute MW information as well as column fractionation method using Mw/Mn measurements. 
MW distribution analysis of SRM706
   
     
      | Analysis method  | 
      Mw x 104 | 
      Mn x 104  | 
      Mw/Mn | 
      
     
      |  
         KF-406LHQ x 2 
        (4.6mmID*250mm)  | 
      26.22 | 
      12.26 | 
      2.13 | 
      
     
      | 26.46 | 
      12.43 | 
      2.13 | 
      
     
      |  
         KF-806M  
        (8.0mmID*300mm)  | 
      26.00 | 
      12.40 | 
      2.10 | 
      
     
      | 28.00 | 
      12.90 | 
      2.17 | 
      
     
      | Light scattering  | 
      25.78 | 
      - | 
      - | 
      
     
      | Sedimentation equilibrum  | 
      28.81 | 
      - | 
      - | 
      
     
      | Column fractionation  | 
      
  | 
      
  | 
      2.10 |