Dynamic Range of Column
Monovalent and divalent cation standards were analyzed using IC 
YK-421. 
This data shows the change in dynamic range of the column with the height of the 
peaks. There is the linearity in the 1 to 200mg/L range for Na
+ 
and K
+ and in the 1 to 100mg/L range for Ca
2+ 
and Mg
2+.

Sample : 
Ca2+
K+
Mg2+
Na+
Column       : Shodex IC YK-421 (4.6mmID*125mm)
Eluent       : 5mM Tartaric acid + 1mM Dipicolinic acid + 1.5g/L Boric acid aq.
Flow rate    : 1.0mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40deg-C