Cation Analysis in the Presence of PEG (YS-50)
Waste water from electronic production factories may contain hydrophilic polymers. As a waste water model, sample containing mono- and di-valent cations in the presence of polyethylene glycol (PEG2000) was prepared. Ion chromatography (cation) column, IC
YS-50 was used under non-suppressor method. All cations analyzed showed good recovery without having any effect from polyethylene glycol. Some PEG2000 is adsorbed on the column, however this can be removed by adding 50% ACN in the eluent. 
.

Sample :10micro-L
Cation mixture in 5% PEG2000 solution
  
    No.  | 
    cation  | 
    Recovery (%)   | 
  
  
    | 1. | 
    Li+ 2mg/L | 
    106  | 
  
  
    | 2. | 
    Na+ 10mg/L  | 
    104  | 
  
  
    | 3. | 
    NH4+ 10mg/L  | 
    101  | 
  
  
    | 4. | 
    K+ 20mg/L  | 
    101  | 
  
  
    | 5. | 
    Mg2+ 10mg/L  | 
    99  | 
  
  
    | 6. | 
    Ca2+ 10mg/L  | 
    97  | 
  
Column       : Shodex IC YS-50 (4.6mmID*125mm) 
Eluent       : 4mM Methanesulfonic acid aq.
Flow rate    : 1.0mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40deg-C