Features of Multimode Columns (Peptide Analysis)
Although Asahipak
GS-HQ series is basically SEC columns,
they permit different types of separation and analysis from conventional ones
by use of a multimode which combines the hydrophobic interaction (RPC) and ionic
interaction (IEX: ion exclusion). The
GS-320 HQ column
performs well in separating hydrophillic peptides, particularly acidic and basic
peptides. As the columns can be used over a wide range of eluent pH because of
being polymer-based column, they can be washed with alkaline solution leading
to a longer column life, which is another feature of the column.
Sample : 20micro-L
1.
Glu-Ala-Glu 0.025%
2.
Arg-Asp 0.05%
3.
Gly-His-Lys 0.025%
4.
Arg-Pro-Lys-Pro 0.025%
|
MW |
pI |
Σf |
Glu-Ala-Glu |
347 |
3.12 |
0.39 |
Arg-Asp |
289 |
6.75 |
0.68 |
Gly-His-Lys |
340 |
9.95 |
0.29 |
Arg-Pro-Lys-Pro |
497 |
11.44 |
3.24 |
pI : Isoelectric point
Σf : Hydrophobic parameter
Column : Shodex Asahipak GS-320 HQ (7.5mmID*300mm)
Eluent : 50mM Ammonium acetate buffer(pH6.7)
Flow rate : 0.5mL/min
Detector : UV(220nm)
Column temp. : 30deg-C