Standard Proteins (18) (KW-803)
When a sample contains high molecular weight portions, PROTEIN
KW-803 (a column for aqueous SEC (GFC)) will
give a better result than Company A column.
Sample : Gel Filtration Standard (Bio Rad)
Peak |
Protein |
PROTEIN KW-803 |
Company A column |
Rt(min) |
Width(min) |
Rs |
Rt(min) |
Width(min) |
Rs |
1 |
Thyroglobulin |
6.5657 |
0.7550 |
|
6.5657 |
0.7550 |
|
|
(MW: 670,000) |
|
|
2.59 |
|
|
2.27 |
2 |
gamma-Globulin |
8.2957 |
0.5825 |
|
7.7203 |
0.8450 |
|
|
(MW: 158,000) |
|
|
2.33 |
|
|
1.84 |
3 |
Ovalbumin |
9.5547 |
0.5000 |
|
8.9840 |
0.5250 |
|
|
(MW: 44,000) |
|
|
1.93 |
|
|
2.39 |
4 |
Myoglobin |
10.4380 |
0.4175 |
|
10.1290 |
0.4350 |
|
|
(MW: 17,000) |
|
|
4.69 |
|
|
3.57 |
5 |
Vitamin B12 |
12.3423 |
0.3950 |
|
11.5647 |
0.3700 |
|
|
(MW: 1,350) |
|
|
|
|
|
|
Columns : (Left); Shodex PROTEIN KW-803 (8.0mmID*300mm), (Right); Company A column
Eluent : 50mM Sodium phosphate buffer(pH7.0) + 0.3M NaCl
Flow rate : 1.0mL/min
Detector : UV(220nm)
Column temp. : Room temp.