Opioid Peptides (2) (By Pre-column Reaction Method)
Opioid peptides were analyzed using Asahipak
ODP-50 6D by pre-column reaction method.
Sample :
1.
Tyr-Gly-Gly-Phe
2.
ME, Met-Enkephalin
3.
LE, Leu-Enkephalin
4.
LE-Arg
5.
alpha-EDP, alpha-Endorphin
6.
LE-NH2
7. DNP 1-7
8. [D-Ala
2] LE-NH
2
9.
ME-Arg-Gly-Leu
10.
gamma-EDP
11.
ME-Arg-Phe
12. DNP 1-8
13.
beta-NED, beta-Neoendorphin
14.
alpha-NED, beta-NED-Lys
15. DNP 1-9
16. DNP 1-10
17. DNP-B
Column : Shodex Asahipak ODP-50 6D (6.0mmID*150mm)
Eluent : (A); 0.1M Borate buffer(pH9.5)/H2O/CH3CN=10/39/1
(B); 0.1M Borate buffer(pH9.5)/H2O/CH3CN=2/3/5
Gradient:(A) to (B)
Flow rate : (0 to 25min); 0.8mL/min, (25 to 55min); 1.0mL/min, (after 55min); 0.8mL/min
Detector : Fluorescence(Ex. 330nm, Em. 430nm)
Column temp. : Room temp.
Guo-Qing Zhang et al., Chem. Pharm. Bull.,
39 (1991) 126.