The study of Injection Volume ; GS-320 Series Columns

The figure shows the effect of the injection volume to the separation for the analysis of uracil using GS-320 series columns.

(Recommended injection volume)
Product
name
Size (mm)
ID x L
Injection volume
(micro-L)
GS320A-1E

1.0 x 250

1
GS320A-2E
2.0 x 250
10
GS-320 HQ
7.5 x 300
100

Sample : Uracil
2ng (GS320A-1E)
10ng (GS320A-2E)
200ng (GS-320 HQ)

Columns  : Shodex GS320A-1E (1.0mmID*250mm)
           Shodex GS320A-2E (2.0mmID*250mm)
           Shodex Asahipak GS-320 HQ (7.5mmID*300mm)
Eluent   : 10mM HCOONH4 buffer(pH4.0)
Detector : UV(260nm) cell volume; 3micro-L, optical path length; 7mm
Tubing   : 0.1mmID PEEK tube
Injector : Rheodyne 8125