Comparison of NH2P with Silica-based Amino Column (1)
Polymer-based Asahipak
NH2P-50 4E column has resolved the
problem occurring with silica-based amino columns, whose performance falls with
the time.
1) Conventional analytical methods used for silica-based amino columns can be
applied to the NH2P column.
Eluent ---- Mixture of water and acetonitrile which is commonly used for silica-based
amino columns can be also used.
Column temperature ---- Can be analyzed at room temperature
2) Excellent quantitative analysis can be performed using NH2P column.
3) Further more, the NH2P column can be washed with alkaline solvent.
Sample :
1.
Fructose
2.
Glucose
3.
Sucrose
4.
Maltose
Columns : Shodex Asahipak NH2P-50 4E, Amino column from other manufacturer (4.6mmID*250mm each)
Eluent : CH3CN/H2O=75/25
Flow rate : 1.0mL/min
Detector : Shodex RI
Column temp. : 30deg-C