Effect of Sample Solvent Composition
Effect of sample solvent composition when Asahipak
NH2P-50
is used is shown. When making sample solution to be injected, sample should be dissolved in a solvent whose composition is
as close as the eluent, preferably in the eluent. When 75% acetonitrile solvent is used for the eluent and a sample cannot be dissolved in it, first,
the sample should be dissolved in a solvent which dissolves the sample, and then, adjust the solvent composition
to be acetonitrile aqueous solution of 50% or more.
Sample : 1mg/mL each, 20micro-L
1.
meso-Erythritol
2.
Xylitol
3.
Fructose
4.
Glucose
5.
Sucrose
6.
Lactose
7.
Maltose
Column : Shodex Asahipak NH2P-50 4E (4.6mmID*250mm)
Eluent : CH3CN/H2O=75/25
Flow rate : 1.0mL/min
Detector : Shodex RI
Column temp. : 25deg-C