Analysis of Soft Drinks (LC/MS)
Canned soft drink containing coffee and sugar, tea and sugar and orange juice were analyzed using LC/MS. Stable baselines were obtained by using a polymer-based amino column, Asahipak
NH2P-50 2D.
Sample :
Soft drink
Canned coffee with sugar
1.
Sucrose(m/z=378)
Canned tea with sugar
1.
Sucrose(m/z=378)
Canned
orange juice
1.
Fructose(m/z=216)
2.
Glucose(m/z=216)
(HPLC)
Instrument : Agilent 1100
Column : Shodex Asahipak NH2P-50 2D (2.0mmID*150mm)
Eluent : H2O/CH3CN=25/75
Flow rate : 0.2mL/min
Column temp. : 40deg-C
(Post column)
Reagent : CH3CN/CHCl3=50/50
Flow rate : 0.2mL/min
(MS)
Instrument : Agilent 1100MSD
Mass range : 100-500(m/z)
Ionization : APCl
Mode : SIM(M+Cl)
Polarity : Negative
Fragmentor : 20V
Nebulizer : N2(40psi)
Drying gas : N2(10L/min, 350deg-C)
Corona current : 30micro-A
Vaporizer temp. : 400deg-C