Analysis of myo-Inositol according to EP Method (SC1011-7F)
According to EP method (Reference No. 1805, Version Date 01/2008), mannitol should be
analyzed using a column which can separate myo-Inositol and mannitol (impurity A) with resolution
of equal to or greater than 4.0. EP
SC1011-7F is a column
specially designed for myo-Inositol analysis, whcih satisfies this criterion.
Sample :
1.
myo-Inositol
2.
Mannitol
3.
Glycerol
Column : Shodex EP SC1011-7F (7.8mmID*300mm)
Eluent : H2O
Flow rate : 0.5mL/min
Detector : Shodex RI
Column temp. : 85deg-C