Analysis of Organic Acids (Eluent Conteinig Cyclodextrin)
Organic acid standards were analyzed using RSpak
KC-811.
The elution time of organic acids which have long carbon chain such as butyric
acid or valeric acid can be shortened adding cyclodextrin in the eluent.
Sample : 5mM each , 20micro-L
1.
Fumaric acid
2.
Succinic acid
3.
Lactic acid
4.
Formic acid
5.
Acetic acid
6.
Propionic acid
7.
n-Butyric acid
8.
Isobutyric acid
9.
n-Valeric acid
10.
Isovaleric acid
Columns : Shodex RSpak KC-LG (8.0mmID*50mm) + KC-811 (8.0mmID*300mm) x 2
Eluent : 2mM HClO4 + 4mM alph-cyclodextrin(pH3.0)
Reagent : 10-fold diluted ST3-R (for post column method)
Flow rate : (Eluent); 1.0mL/min, (Reagent); 0.5mL/min
Detector : VIS(430nm)
Column temp. : 47deg-C