Basic Drugs : Effect of Ion Concentration
Effect of ion concentration on elution pattern for the analysis of scopolamine
which has tertiary amine and whose pKa is 7.6 is shown. When the ion concentration
is lowered, non-specific adsorption is observed for ODS-A and sample is completely
adsorbed at the ion concentration lower than 10mM. This tendency is also observed
even for ODS-C which is a polymer-coated ODS column and sample is completely adsorbed
at 0mM. To the contrary, Asahipak
ODP columns are free
from the non-specific adsorption problem at any ion concentration level. Then
ODP columns are suitable for analysis with the ESI method (LC/MS) in which ionization
of sample is often supressed under high-salt concentration.
Sample : 1.0mg/mL
Scopolamine, 50micro-L
Columns : Shodex Asahipak ODP-50 4D (4.6mmID*150mm), ODS-A, ODS-C (4.6mmID*150mm each)
Eluent : Phosphate buffer(pH7)/CH3CN=60/40
Flow rate : 0.6mL/min
Detector : UV(254nm)
Column temp. : 30deg-C