KF-801 | KF-801 with the smallest pore size can separate EPIKOTE 828 which has the smallest molecular weight. However, this column cannot separate EPIKOTE 1001, 1004 and 1007 well because their molecular weight exceed the exclusion limit of KF-801. |
KF-802 | KF-802 can separate EPIKOTE 828 and 1001, but it cannot separate EPIKOTE 1004 and 1007 well because the molecular weight is above the exclusion limit. |
KF-803 | KF-803 can separate all EPIKOTE 828, 1001, 1004 and 1007. However the separation of EPIKOTE 828 is not as good as those of KF-801 and 802 because the pore size of KF-803 is too large for EPIKOTE 828. |
KF-804 | KF-804 can also separate all EPIKOTE 828, 1001, 1004 and 1007. But none of the separations are as good as those of KF-803 because the pore size of KF-804 is too large for all EPIKOTE. KF-804 is suited for the epoxy resins, whose molecular weight is Larger than EPIKOTE 1007. |