Epoxy Resin (1) (Comparison of KF-803L and KF-603 + KF-602.5)

Two kinds of epoxy resin, epokote 1001 and epikote 1004 were analysed using conventional type organic SEC (GPC) column (GPC KF-803L) and downsized organic SEC (GPC) columns (GPC KF-603 + KF-602.5). Using downsized GPC columns, better separation could be obtained in shorter time. For downsized GPC columns, an RI detector of small cell volume type was used.



Sample : Epikote (Epoxy reisne)
Columns      : Shodex GPC KF-803L (8.0mmID*300mm) x 2   
Eluent       : THF                                     
Flow rate    : 1.0mL/min                                
Detector     : UV(254nm)(conventional type)                         
Column temp. : Room temp.
 
Columns      : Shodex GPC KF-603 + KF-602.5 (6.0mmID*150mm each)  
Eluent       : THF                             
Flow rate    : 0.6mL/min
Detector     : Shodex RI(small cell volume) 
Column temp. : 40deg-C