Selection of Column (1) (Sodium Poly(Styrene Sulfonate))
Calibration curves of Asahipak
GF-7M HQ (packed with poly(vinyl
alcohol)) and OHpak
SB-806M HQ (packed with poly(hydroxy
methacrylate)) are compared using a hydrophilic polymer standerd, sodium poly(styrene
sulfonate). The comparison shows that the calibration curve of Asahipak GF-7M
HQ is more linear than that of OHpak SB-806M HQ.
Sample :
Sodium poly(styrene sulfonate) standard
Columns : Shodex Asahipak GF-7M HQ (7.5mmID*300mm), OHpak SB-806M HQ (8.0mmID*300mm)
Eluent : 50mM Na2HPO4 aq./CH3CN=90/10
Flow rate : 0.6mL/min
Detector : Shodex RI
Column temp. : 40deg-C