Polyacrylamide (2) (KS-807, SB-806 HQ)
The analytical results of polyacrylamide are shown. In the analysis with OHpak
SB-806 HQ (exclusion limit: 2.0x10
7g/mol),
polyacrylamide elutes around exclusion limit. On the other hand, in the analysis
with SUGAR
KS-807 (exclusion limit: 2.0x10
8g/mol),
polyacrylamide elutes with SEC mode.
Sample : 100micro-L
0.02%
Polyacrylamide (PAAm)
(Scientific Polymer Product)
Columns : Shodex SUGAR KS-807, OHpak SB-806 HQ (8.0mmID*300mm each)
Eluent : 0.2M NaCl aq.
Flow rate : 0.5mL/min
Detector : Shodex RI and MALS(Multi angle laser light scattering detector)
Column temp. : 30deg-C