Polyacrylamide (2) (KS-807, SB-806 HQ)

The analytical results of polyacrylamide are shown. In the analysis with OHpak SB-806 HQ (exclusion limit: 2.0x107g/mol), polyacrylamide elutes around exclusion limit. On the other hand, in the analysis with SUGAR KS-807 (exclusion limit: 2.0x108g/mol), polyacrylamide elutes with SEC mode.


Sample : 100micro-L
0.02% Polyacrylamide (PAAm)
(Scientific Polymer Product)

Columns      : Shodex SUGAR KS-807, OHpak SB-806 HQ (8.0mmID*300mm each)
Eluent       : 0.2M NaCl aq.
Flow rate    : 0.5mL/min
Detector     : Shodex RI and MALS(Multi angle laser light scattering detector)
Column temp. : 30deg-C