The Effect of The Unit of Columns on The Separation (1) (LF-404: PS)
Each one, two and three units of GPC
LF-404, were used for the analysis of
the mixture of standard polystyrene compounds. Only a single unit of LF-404 can determine
a broad range of molecular weights from 300 to 2,000,000. However the resolution is
higher depending on the number of columns, so it is better to use several units of columns
for obtaining more accurate molecular weight distribution.
Sample : 10micro-L
1.
PS, Polystyrene (MW: 1,030,000)
2.
PS, Polystyrene (MW: 152,000)
3.
PS, Polystyrene (MW: 66,000)
4.
PS, Polystyrene (MW: 22,000)
5.
PS, Polystyrene (MW: 5,050)
6.
PS, Polystyrene (MW: 580)
7.
Ethylbenzene
Columns : Shodex GPC LF-404 (4.6mmID*250mm) x n
Eluent : THF
Flow rate : 0.3mL/min
Detector : Shodex RI(small cell volume)
Column temp. : 40deg-C