Transition Metal Ions (2) (T-521)
Transition metal ions were analyzed using IC
T-521. An oxalic acid/tartaric acid eluent is used for the analysis of transition
metal ions. Each transition metal ion generates complex ions with the organic acids in the eluent at a different rate. The easier the
transition metal ions generate the complex ions, the faster they elute. The sample should be dissolved in a 40mM HCl aqueous solution.
Sample : 70micro-L
1.
Fe3+
2.
Cu2+
3.
Ni2+
4.
Zn2+
5.
Co2+
6.
Pb2+
7.
Fe2+
Column : Shodex IC T-521 (4.6mm*150mm)
Eluent : 6mM Oxalic acid + 3mM Citric acid + 0.1mM NaN3 + 0.1mM NaHSO3(pH4.1 adjusted by KOH)
Reagent : 0.2mM *PAR + 3M NH4OH + 1M CH3COOH (Mixing coil(0.5mmID*1m))
Flow rate : (Eluent); 1.0mL/min, (Reagent); 1.0mL/min
Detector : VIS(530nm)
Column temp. : 40deg-C
*PAR : 4-(2-Pyridylazo)resorcinol