Tap Water (2) (T-521)
Transition metal ions in tap water were analyzed using IC
T-521. Fe
3+ and Zn
2+ are
contained in tap water. HCl is added to the sample to ionize Fe to Fe
3+ completely. (The final HCl concentration is 40mM)
Sample :
Tap water
1.
Fe3+
2.
Zn2+
Column : Shodex IC T-521 (4.6mm*150mm)
Eluent : 6mM Oxalic acid + 3mM Citric acid + 0.1mM NaN3 + 0.1mM NaHSO3(pH4.1 adjusted by KOH)
Reagent : 0.2mM *PAR + 3M NH4OH + 1M CH3COOH (Mixing coil(0.5mmID*1m))
Flow rate : (Eluent); 1.0mL/min, (Reagent); 1.0mL/min
Detector : VIS(530nm)
Column temp. : 40deg-C
*PAR : 4-(2-Pyridylazo)resorcinol