Effect of Organic Solvents
For the analysis of nonylphenol ethoxylate using Asahipak 
GF-310 
HQ (a multisolvent GPC column), most of organic solvents can be used.
The figure shows the elution characteristics of nonylphenol ethoxylate when typical 
organic solvents are used. The elution time of nonylphenol ethoxylate is the fastest 
when n-propanol is used, medium when acetonitrile is used and the latest when 
methanol is used. The order is as same as the case when reversed phase columns 
are used.

Sample : 
Nonylphenol ethoxylate (n=20)
Column       : Shodex Asahipak GF-310 HQ (7.5mmID*300mm)
Flow rate    : 0.6mL/min
Detector     : UV(280nm)
Column temp. : 30deg-C