Effect of Temperature
The effect of temperature when POE distribution of octylphenyl ethoxylate (n=6) 
is measured using Asahipak 
GF-310 HQ ( a multisolvent GPC 
column) is shown here. Over the wide temperature range, good chromatograms are 
obtained. At high temperature, higher resolution is obtained because of sharp 
peak shape though the retention time is faster. 
Sample : 
Octylphenyl ethoxylate
Column       : Shodex Asahipak GF-310 HQ (7.5mmID*300mm)
Flow rate    : 0.6mL/min
Eluent       : CH3CN/H2O=60/40
Detector     : UV(280nm)
Column temp. : 15, 30, 60deg-C