Agricultural Chemicals (1)
The analyses of agricultural chemicals using two columns, RSpak
DE-613 and
GOLF-413 were performed. The analysis time can be shortened by using RSpak GOLF-413.
Sample :
1.
Asulam, 2.
Oxine-Cu,
MCPP;
2. (Methyl-4-chlorophenoxy)-propionic acid
3.
Thiuram, 4.
Iprodione ,5.
Bensulide,
Pencycuron
Column : Shodex RSpak DE-613 (6.0mmID*150mm)
Eluent : 50mM Sodium phosphate buffer(pH3.3)/CH3CN=49/51
Flow rate : 1.0mL/min
Detector : UV(0min to 6.9min to 17.0min, 270nm to 243nm to 230nm)
Column temp. : 40deg-C
Column : Shodex RSpak GOLF-413 (4.6mmID*150mm)
Eluent : 50mM Sodium phosphate buffer(pH3.2)/CH3CN=48/52
Flow rate : 1.0mL/min
Detector : UV(0min to 3.5min to 5.0min, 270nm to 243nm to 230nm)
Column temp. : 40deg-C