Detection Limit of Cyanide
Combining IEC
DEAE-825 (a column for weak anion exchange chromatography) with an electrochemical detector
(ECD) has permitted sensitive detection of cyanide. With this method, the detection
limit of cyanide is 0.2 micro-g/L.
Sample :
CN-
Column : Shodex IEC DEAE-825 (8.0mmID*75mm)
Eluent : (10mM Sodium carbonate + 1mM Ethylenediamine) aq./CH3OH=90/10
Flow rate : 1.0mL/min
Detector : ECD(Electrode; Ag 0mV SCE)
Column temp. : 40deg-C