Arsenic Compounds (1)
RSpak
NN-614 is packed with poly(hydroxymethacrylate) bonded
with trace amounts of sulfo groups, and the main separation mode is the reversed
phase mode. However, due to the sulfo groups, ion interaction also works. Acidic
substances are separated by ion exclusion mode and elute faster. Neutral substances
are separated by reversed phase and alkaline substances are separated by a mixture
of reversed phase (in the case of hydrophobic structure) and ion exchange mode.
Therefore, the elution order is basically as follows: first the acidic, then the
neutral and finally the alkaline substances. The column is suited to the analysis
of mixture of acidic, neutral and alkaline substances.
Sample :
Arsenic compounds,(Above) 100ppm each, (Below)
10ppb each, 20micro-L
Monomethylarsonic acid, MMAA
Dimethylarsinic acid, DMAA
Arsenobetaine, AB
Tetramethylarsonium, TeMA
Trimethylarsine oxide, TMAO
(HPLC)
Instrument : Agilent 1100
Column : Shodex RSpak NN-614 (6.0mmID*150mm)
Eluent : 8mM HNO3/5mM NH4NO3 aq.
Flow rate : 0.4mL/min
Column temp. : 40deg-C
(MS)
Instrument : Agilent 1100MSD
Mass range : (Above): TIC 100-1000(m/z)
(Below): SIM
Ionization : Electrospray
Polarity : Positive
Fragmentor : 80V
Nebulizer : N2(60psi)
Drying gas : N2(10L/min, 350deg-C)