Effects of Temperature on GPC Clean-up
Two different temperature settings (room temerature and 40 deg-C) were compared for the GPC clean-up method using CLNpak
EV-2000 AC. At room temperature, the high viscosity causes peak broadening and increases the analysis time. Therefore, analysis at 40 deg-C is preferred. In Food Sanitation Law #1003001 (Released by Ministry of Health, labour and welfare in Japan, 2006), analysis temperature is set at 40 deg-C.
Sample : 5mL
Fatty tuna
Acrinathrin 40mg/L
Tricyclazole 40mg/L
Column : Shodex CLNpak EV-G AC (20.0mmID*100mm) + EV-2000 AC (20.0mmID*300mm)
Eluent : Acetone/Cyclohexane=1/4
Flow rate : 5mL/min
Detector : UV(254nm)