Tryptic Digest of Human IgG
When a tryptic digest of human lgG was analyzed using PROTEIN
KW-803
and
KW-802.5 (columns for Aqueous SEC (GFC)) with a UV detector
(230nm) as well as with an optical rotation detector (ORD). The ORD shows a negative
peak and it gives the information that human lgG is levorotatory.

Sample : Triptic digest of
human IgG
Columns : Shodex PROTEIN KW-803 + KW-802.5 (8.0mmID*300mm each)
Eluent : 0.05M Tris-HCl buffer(pH7.6) + 0.15M NaCl + 0.3mM CaCl2
Flow rate : 0.5mL/min
Detector : UV(230nm), ORD
Column temp. : Room temp.