Comparison of Analysis of Organic Acids with 4 Kinds of Detectors (Soy Sauce)
Comparison of Analysis of Organic Acids using RSpak
KC-811
with 4 Kinds of Detectors was shown.
When analyzing organic
acid satndards, 4 kinds of detectors are usable. However, when analyzing actual sample, we recommend to use the post-column method
which allows high-selective analysis of organic acids. For example, soy sauce
is a complicated mixture of organic acids, amino acids, saccharides, NaCl, ions
and organic compounds. With UV detector (210nm), too many organic substances,
organic compounds and organic acids are detected. With a conductivity detector,
organic acids cannot be quantified satisfactory, as the NaCl peak is appearing
at around 10 minutes is large. Also it is difficult to quantify organic acids
with an RI detector, as almost all the components including saccharides are detected.
Sample :
Soy sauce
1.
Citric acid
2.
Pyruvic acid
3.
Malic acid
4.
Succinic acid
5.
Lactic acid
6.
Formic acid
7.
Acetic acid
8.
Pyroglutamic acid
Columns : Shodex RSpak KC-G (6.0mmID*50mm) + KC-811 (8.0mmID*300mm) x 2
Eluent : 3mM HClO4 aq.
Flow rate : 1.0mL/min
Column temp. : 50deg-C