Selection of Column (3) (Poly(Sodium Methacrylate))
Calibration curves of Asahipak
GF-7M HQ (packed with poly(vinyl alcohol)) and OHpak
SB-806M HQ
(packed with poly(hydroxy methacrylate)) are compared using a hydrophilic polymer standerd, poly(sodium methacrylate).
The comparison shows that the calibration curve of Asahipak GF-7M HQ is more linear than that of OHpak SB-806M HQ.

Sample :
Poly(sodium methacrylate) standard
Columns : Shodex Asahipak GF-7M HQ (7.5mmID*300mm), OHpak SB-806M HQ (8.0mmID*300mm)
Eluent : 50mM Na2HPO4 aq./CH3CN=90/10
Flow rate : 0.6mL/min
Detector : Shodex RI
Column temp. : 40deg-C