Anion standards were analyzed using IC I-524A. This eluent is suitable for the separation of ions that elute before chloride. After each sample is analyzed, it is necessary to wash the column with 200 to 300micro-L of 0.1M tartaric acid when the sample contains NO3-, SO42-, etc. because they elute very slowly and may be remaining inside the column.


Sample : 10mg/L each, 50micro-L
1. Acetic acid
2. H2PO4-
3. H2PO2-
4. H2PO3-
5. ClO2-
6. BrO3-
7. Cl-

Column       : Shodex IC I-524A (4.6mmID*100mm)
Eluent       : 1mM Tartaric acid aq.
Flow rate    : 1.2mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40deg-C