River Water (1) (I-524A)

Anions in river water were analyzed using IC I-524A. In river water, small volume of H2PO4-, Br- and NO2- as well as Cl-, NO3- and SO42- are present. This eluent can be used to separate H2PO4- and F-.


Sample : 100micro-L each
Anion standards
1. H2PO4- 10mg/L
2. F- 1mg/L
3. Cl- 1mg/L
4. NO2- 5mg/L
5. Br- 5mg/L
6. NO3- 5mg/L
7. SO42- 5mg/L

5-fold diluted
River water (Water)
1. H2PO4-
3. Cl-
4. NO2-
5. Br-
6. NO3-
7. SO42-

Column       : Shodex IC I-524A
Eluent       : 2.0mM Phthalic acid + 1.84mM Tris(hydroxymethyl)aminomethane + 300mM Boric acid aq.
Flow rate    : 1.2mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40deg-C