Cation Analysis in the Presence of PEG (YS-50)

Waste water from electronic production factories may contain hydrophilic polymers. As a waste water model, sample containing mono- and di-valent cations in the presence of polyethylene glycol (PEG2000) was prepared. Ion chromatography (cation) column, ICYS-50 was used under non-suppressor method. All cations analyzed showed good recovery without having any effect from polyethylene glycol. Some PEG2000 is adsorbed on the column, however this can be removed by adding 50% ACN in the eluent. .


Sample :10micro-L
Cation mixture in 5% PEG2000 solution
No.
cation
Recovery (%)
1. Li+ 2mg/L
106
2. Na+ 10mg/L
104
3. NH4+ 10mg/L
101
4. K+ 20mg/L
101
5. Mg2+ 10mg/L
99
6. Ca2+ 10mg/L
97


Column       : Shodex IC YS-50 (4.6mmID*125mm) 
Eluent       : 4mM Methanesulfonic acid aq.
Flow rate    : 1.0mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40deg-C