Anion Standards (19) (SI-90 4E)
Please use this column with suppressor type ion chromatography system.
Anion standards were analyzed using IC
SI-90 4E (a column
for anions analysis). For general purpose, it is recommended to use 1.8mM Na
2CO
3
+ 1.7mM NaHCO
3 as the eluent. This eluent is best suited
to the separation of typical 7 anions, F
-,Cl
-,NO
2-,Br
-,NO
3-,PO
43-,SO
42-.
Using the flow rate of 1.5 mL/min, seven anions can be analyzed within ten minutes.
Use the column temperature of 25deg-C.
Sample : 20micro-L
1.
F- (2ppm),
2.
Cl- (3ppm),
3.
NO2- (5ppm),
4.
Br- (10ppm)
5.
NO3- (10ppm),
6.
HCO3- (300ppm),
7.
PO43- (15ppm),
8.
SO42- (15ppm)
(How to prepare the eluent)
(1) Put 19.08g of reagent grade Na
2CO
3 and 14.28g of reagent grade
NaHCO
3 into a 1 liter messflask.
(2) Dilute with purified water to make up a 1 liter solution.
(3) Dissolve the solution using an ultrasonic vibrator. When the solution volume decrease after the vibration, add pure water to make up exact 1 liter solution. This is the concentrated solution and can be stored for half year in a refregirator with stopper.
(4) When making the eluent, take 10ml of the concentrated solution and dilute with purified water to make up 1 liter solution. Pass the eluent through a 0.45micron membrane filter before use.
Column : Shodex IC SI-90 4E (4.0mmID*250mm)
Eluent : 1.8mM Na2CO3+ 1.7mM NaHCO3 aq.
Flow rate : 1.0mL/min or 1.5mL/min
Detector : Suppressed conductivity
Column temp. : Room temp. (25deg-C)