Bromate in Drinking Water
Please use this column with suppressor type ion chromatography
system.
Recently ozonation has been applied for sterilization of drinking water. Ozonation
transforms part of bromide to bromate which is a potential carcinogen. So, it
is very important to measure bromate in drinking water. A modeldrinking water
with bromate was measured by suppressed coductivity with IC
SI-50
4E.
Sample : Model drinking
Water, 500micro-L
1. 0.1mg/L
F-
2. 0.003mg/L
BrO3-
3. 20mg/L
Cl-
4. 0.1mg/L
Br-
5. 10mg/L
NO3-
6. 10mg/L
SO42-
1. : 0.02 mg/L
2. : 0.01 mg/L
3. : 0.003mg/L
Column : Shodex IC SI-50 4E (4.0mmID*250mm)
Eluent : 3.2mM Na2CO3 + 1mM NaHCO3 aq.
Flow rate : 0.8mL/min
Detector : Suppressed conductivity
Column temp. : 25deg-C