Sea Water (3) (SI-90 4E)
Please use this column with suppressor type
ion chromatography system.
Anions in sea water were analyzed by anion-chromatography column, IC
SI-90 4E Compared to chloride which is highly concentrated in sea water, concentration of nitrous acid is too low, and thus interfered by chloride ion peak to be monitored by a CD. Since nitrous, bromine, and nitrate ions have UV absorbance, they can be monitored by UV detector without being affected by chloride ion.
Sample : 100-fold diluted
Sea water (
Water), 20micro-L
1.
F-
2.
Cl-
3.
NO2-
4.
Br-
5.
NO3-
6.
SO42-
Column : Shodex IC SI-90 4E (4.0mmID*250mm)
Eluent : 1.8mM Na2CO3 + 1.7mM NaHCO3 aq.
Flow rate : 1.0mL/min
Detector : Suppressed conductivity, UV(210nm)
Column temp. : 25deg-C