Table of Anion retention Time under Various Eluent Conditions (SI-90 4E)
Please use this column with suppressor type
ion chromatography system.
Common anions were analyzed with anion analysis column, IC
SI-90 4E, under different eluent conditions. Result is summarized in the table.
(1) Optimal/recommended condition: 1.8 mM Na
2CO
3 + 1.7 mM NaHCO
3 aq.
(2) For sample containing high carbonate concentration or for high sensitivity analysis: 1.0 mM Na
2CO
3 + 4.0 mM NaHCO
3 aq.
(3) Eluent of strongest retention among 3 conditions: 12.0 mM NaHCO
3 aq.
Sample :
F-
Acetic acid
Formic acid
Methacrylic acid
Cl-
NO2-
Br-
NO3-
PO43-
SO42-
Oxalic acid
I-
WO42-
MoO42-
S2O32-
CrO42-
SCN-
Column : Shodex IC SI-90 4E (4.0mmID*250mm)
Eluent : (1) 1.8mM Na2CO3 + 1.7mM NaHCO3 aq.
(2) 1.0mM Na2CO3 + 4.0mM NaHCO3 aq.
(3) 12mM NaHCO3 aq.
Flow rate : 1.0mL/min
Detector : Suppressed conductivity
Column temp. : 25deg-C