Comparison of CARB-413 Column and ODS Column (2)
The comparison of anaysis of N-methylcarbamates standards using RSpak 
CARB-413 
column with isocratic elution and that using ODS column with gradient 
elution.
A good result was obtaioned using CARB-413 column. The time required for one anaylsis 
is almost same for both two cases. But, additional time is necessary required 
for reconditioning the system in the case of gradient elution though such additinal 
time is unnecessary in the case of isocratic elution. As the rerult, total average 
time required for one analysis with isocratic elution is about a half of that 
with gradient elution. 

Sample : 
1. 
Aldicarb sulfone
2. 
Methomyl
3. 
3-Hydroxycarbofuran
4. 
Dioxacarb
5. 
Aldicarb
6. 
Propoxur
7. 
Carbofuran
8. 
Carbaryl
9. 
Premecarb
10. 
Methiocarb
Column           : RSpak CARB-413 (4.6mmID*150mm)
Other conditions : Refer to Analysis Method of N-methyl Carbamates