Hydrolyzed Dextran (5)
The change of hyrolyzed dextran is analyzed. The analysis was done using SUGAR
KS-802.
(Sample preparation)
1) Hydrolyze 4% dextran aquaous solution using 1M HCl at 100 deg-C.
2) Take sample before and after 30, 40 and 50min from the start of the hydrolyzation.
3) Neutralize the sample adding cold 1M NaOH aaquaous solution.
4) Desalt the sample passing it through ion-exchange resin.
5) Filtrate with 0.45 micron membrane filter.
Sample :
Hydrolyzed dextran, 20micro-L (
Dextran)
Columns : Shodex SUGAR KS-802(8.0mmID*300mm) x 2
Eluent : H2O
Flow rate : 1.0mL/min
Detector : Shodex RI
Column temp. : 80deg-C