Anion Standards (1) (I-524A)
Anion standards were analyzed using IC
I-524A. A phthalic
acid eluent is recommended for the general analysis. Most of the important
anions such as F
-or H
2PO
4-,
Cl
-, NO
2-, Br
-,
NO
3- and SO
42-
can be analyzed with this eluent. Therefore, it is suitable for the routine analysis
of anions.
Sample : 20micro-L
1.
H2PO4- 30mg/L
2.
Cl- 5mg/L
3.
NO2- 10mg/L
4.
Br- 10mg/L
5.
NO3- 10mg/L
6.
SO42- 10mg/L
Column : Shodex IC I-524A (4.6mmID*100mm)
Eluent : 2.5mM Phthalic acid + 2.3mM Tris(hydroxymethyl)aminomethane aq.
Flow rate : 1.2mL/min
Detector : Non-suppressed coductivity
Column temp. : 40deg-C