Effect of Sample Load on Column Efficiency (1) (I-524A)
Anion standards were analyzed using IC
I-524A. This shows
the change in column efficiency with the concentration of the sample when the
injection volume is 100micro-L. The column efficiency does not decrease until
the sample concentration reaches 10mg/L for Cl
- and
100mg/L for SO
42-. Beyond those concentrations,
HETP increases and peak becomes broad. When 1mg/L of Cl
-
is injected, there is not decreasing in column efficiency until the injection
volume reaches 500micro-L.
Sample :
Cl-
SO42-
HETP=(Column length)/(Theoretical plate number)
Column : Shodex IC I-524A (4.6mmID*100mm)
Eluent : 2.5mM Phthalic acid(pH4.0 adjusted by Tris(hydroxymethyl)aminomethane)
Flow rate : 1.2mL/min
Detector : Non-suppressed coductivity
Column temp. : 40deg-C