Anion Standards (6) (I-524A)

Anion standards were analyzed using IC I-524A. By adding boric acid to the phthalic acid, phosphoric ion (H2PO4-) and fluoride ion (F-) can be separated.


Sample : 100micro-L
1. H2PO4- 10mg/L
2. F- 1mg/L
3. Cl- 1mg/L
4. NO2- 5mg/L
5. Br- 5mg/L
6. NO3- 5mg/L
7. SO42- 5mg/L

Column       : Shodex IC I-524A (4.6mmID*100mm)
Eluent       : 2.0mM Phthalic acid + 1.84mM Tris(hydroxymethyl)aminomethane + 300mM Boric acid aq.
Flow rate    : 1.2mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40deg-C