River Water (1) (I-524A)
Anions in river water were analyzed using IC
I-524A. In river water, small volume of H
2PO
4-, Br
- and NO
2- as well as Cl
-, NO
3- and SO
42- are present. This eluent can be used to separate H
2PO
4- and F
-.
Sample : 100micro-L each
Anion standards
1.
H2PO4- 10mg/L
2.
F- 1mg/L
3.
Cl- 1mg/L
4.
NO2- 5mg/L
5.
Br- 5mg/L
6.
NO3- 5mg/L
7.
SO42- 5mg/L
5-fold diluted
River water (
Water)
1.
H2PO4-
3.
Cl-
4.
NO2-
5.
Br-
6.
NO3-
7.
SO42-
Column : Shodex IC I-524A
Eluent : 2.0mM Phthalic acid + 1.84mM Tris(hydroxymethyl)aminomethane + 300mM Boric acid aq.
Flow rate : 1.2mL/min
Detector : Non-suppressed coductivity
Column temp. : 40deg-C