Sea Water (1) (I-524A)

Anions in sea water were analyzed using IC I-524A. Since sea water contains a considerable amount of chloride ion, the chlorine should be removed prior to analysis. Put Ag-type cation exchanger in a disposable syringe and pass the sea water through the syringe. After the pretreatment, other anions can be analyzed without the interference from chloride ion.


Sample : Sea water (Water)
1.F-
2.SO42-

Column       : Shodex IC I-524A (4.6mmID*100mm)
Eluent       : 2.5mM Phthalic acid(pH4.0 adjusted with Tris(hydroxymethyl)aminomethane)
Flow rate    : 1.2mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40deg-C