Sea Water (1) (I-524A)
Anions in sea water were analyzed using IC
I-524A. Since sea water contains a considerable amount of chloride ion,
the chlorine should be removed prior to analysis. Put Ag-type cation exchanger in a disposable syringe and pass the sea water through
the syringe. After the pretreatment, other anions can be analyzed without the interference from chloride ion.
Sample :
Sea water (
Water)
1.
F-
2.
SO42-
Column : Shodex IC I-524A (4.6mmID*100mm)
Eluent : 2.5mM Phthalic acid(pH4.0 adjusted with Tris(hydroxymethyl)aminomethane)
Flow rate : 1.2mL/min
Detector : Non-suppressed coductivity
Column temp. : 40deg-C